Home / Products / Scientific Instruments / Infrared Tympanic Thermometer Calibration Blackbody 336
Model 336
Infrared Tympanic Thermometer Calibration Blackbody 336
HOTECH
【APPLICATIONS】
Ear / forehead thermometer can be verified and calibrated from Model 336. It offers the precise and stabilized temperature reference that allows the lab / manufacturer performs the service of temperature calibration. Model 336 is specially design to follows with ASTM standard. Internal blackbody structure was uniquely designed to generates high infrared emissivity with the temperature stability of ±0.015°C @ 37°C, Stable and constant temperature is guaranteed; moreover, it can be re-tune to get the best control if it is under the diversity ambient temperature.
HOTECH
Feature
01
Unique design for non-contact infrared ear thermometer calibration.02
Intuitive operation, no special training requires.03
For Hospital, Biomedical department, nursing station, factory, distributor or any situation needs to verify the preciseness of ear thermometer.04
Precise temperature control with high stability.05
Small in size, lightweight and easy to carry.06
Special ambient temperature control for great black-body stabilization.07
Precise temperature control with resolution of 0.001°C.Specification
MODEL | 336 |
---|---|
Temperature Range | 25°C to 45°C |
Temperature Control | Microprocessor PID control (Temp. sensor: PT-100) |
Temperature Display | 0.001℃ |
Stability | ±0.015°C @37℃ |
Heat-up Time | 15 minutes minimum |
Ambient Operating Range | 15°C – 25°C(Air-conditioned environment) |
Ambient Operating stability | ±1℃ |
Cover | 1 pc |
Power | 1 A/110V, 0.5/240V ± 10%,50/60 Hz |
Dimensions Target Size | Aperture:1-Forehead 25mm and 2-Eer:11mm |
Dimensions Overall | W296 x D237 x H165mm |
Weight (Net) | 5.2 kgs |